Measuring the dose, not the time
A new peer-reviewed study in Communications Engineering used Zimmer & Peacock (ZP) wire sensors to monitor cold atmospheric plasma (CAP) treatment in real time — and to switch the plasma off automatically once the right dose was reached.
CAP is used clinically for chronic wounds and diabetic ulcers. Much of its effect comes from reactive oxygen and nitrogen species (RONS), such as hydrogen peroxide (H₂O₂). Too little does nothing; too much damages tissue. Yet treatments are usually dosed by exposure time, which ignores differences between patients, wounds, humidity and device distance.
The researchers set out to measure what the plasma actually delivers. They needed sensors small enough to sit in a 50 µL drop of buffer on a mouse wound, robust enough to survive repeated plasma exposure, and fast enough to feed a controller. They chose ZP's H₂O₂ and oxidation–reduction potential (ORP) wire sensors.
The ZP kit behind the study
Five ZP products are named in the paper's methods.
| Product | Part number | Role in the study |
|---|---|---|
| H₂O₂ wire sensor | ZPS-HPE-003-00260 |
Amperometric H₂O₂ detection at a Pt/Ir electrode with a Nafion selective layer |
| ORP wire sensor | ZPS-ORP-003-00258 |
Open-circuit measurement of the overall redox balance |
| Anapot EIS potentiostat | ZP1000080 |
Data acquisition for the static PBS, cell and wound measurements |
| ORP calibration kit | ZP1001687 |
Ferro/ferricyanide standards for ORP calibration |
| Biosensor rinse solution | ZPCH 900-000-00455 |
Rinsing sensors between measurements |
Both wire sensors use a 70 % platinum / 30 % iridium working electrode, chosen because it keeps its catalytic properties under heavy exposure to plasma-generated species, and an integrated Ag/AgCl reference.
How the sensors performed
The H₂O₂ sensor delivered sub-micromolar detection and stayed linear across the full range the study needed.
| Parameter | H₂O₂ sensor | ORP sensor |
|---|---|---|
| Operating mode | Chronoamperometry at +0.65 V | Open-circuit potentiometry |
| Limit of detection / resolution | 0.5 µM (resolution ≈ 0.1 µM) | ≈ 1 mV |
| Linear / dynamic range | 0.5–200 µM | −500 to +500 mV |
| Sensitivity | 12 nA µM⁻¹ | 59 mV per decade (Nernstian) |
| Repeatability across calibration checks | 5 % RSD | 2 % RSD |
| Response time | 20 ± 10 s | 20 ± 10 s |
The sensors worked in three increasingly demanding settings: plain PBS, live keratinocyte cultures and open wounds in mice. In each, both signals rose with plasma dose. On wounds, the optimal 180 s treatment — the one that gave the smallest scars — corresponded to about 60 µM H₂O₂ and about 100 mV ORP, giving future controllers a measurable target. The authors reported no stability or reusability problems over the study.
Closing the loop: the sensor decides when to stop
The headline result is closed-loop control: the H₂O₂ sensor reading, not a timer, switched the plasma off.
A microcontroller read the H₂O₂ sensor every 20 s while the plasma ran. In PBS, it stopped treatment automatically when the concentration reached a 40 µM target, at 270 s. On mouse wounds, the controller stopped once H₂O₂ had risen 25 µM above each wound's own baseline. That took 186.8 s, 189.8 s and 151.8 s in the three animals — the same delivered dose, reached at different times.
That spread is the point. A fixed treatment time risks under- or over-dosing individual wounds; a sensor-based endpoint adapts to the individual wound, the local humidity and small differences in treatment distance. For the in vivo work, the team ran the H₂O₂ sensor in pulsed chronoamperometry, refreshing the diffusion layer every ~20 s for a cleaner signal during live treatment.
Taking the sensors into your own application
All measurements in the study used neutral PBS. The same sensors can be taken into other plasma-activated liquids and process conditions, provided the method is tuned to them.
- Set the potential for your conditions. +0.65 V was optimised for neutral buffer. H₂O₂ oxidation releases protons, so the optimal potential shifts by roughly 59 mV per pH unit; a quick potential series in your own solution finds the plateau.
- Calibrate in your own matrix. Use standards made up in the same liquid you will measure, not only in PBS.
- Keep the reference stable. The integrated Ag/AgCl reference needs some chloride in the sample to hold a steady potential.
- Check for interferents. Run blanks containing any species your plasma produces in quantity, such as nitrite.
The team behind this paper took the same route: they adapted the sensors to a new measurement environment, and it paid off in a published result.
Read the paper and get the kit
The full open-access article: "Electrochemical sensors for in situ monitoring of reactive species during cold atmospheric plasma-based therapies", Communications Engineering 5, 6 (2026), doi:10.1038/s44172-025-00560-w.
To set up real-time H₂O₂ and ORP monitoring in your own plasma work, ask us for a quotation on the H₂O₂ wire sensor (ZPS-HPE-003-00260), the ORP wire sensor (ZPS-ORP-003-00258) and the Anapot EIS potentiostat (ZP1000080).
This article describes published research and makes no claim of clinical or regulatory approval for ZP products.