Cyclic voltammetry (CV) is the classic first experiment in electrochemistry. The potential of an electrode is swept up and back while the current is recorded, and the result – the voltammogram – shows where a chemical is oxidised and reduced, and how quickly its electrons move.
The single most important setting in a CV is the scan rate: how fast the potential is swept. Change it, and the whole voltammogram changes – the peaks grow, and on some electrodes they move apart. This simulator lets you see exactly how.
It recreates a CV of 5 mM ferricyanide – the standard test solution for electrodes – on our ZP 501 screen-printed carbon electrode. The left panel shows the potential applied over time; the right panel shows the current that results. Both are drawn together, so you can see which part of the sweep produces each part of the voltammogram.
How to use it: press Play the scan to watch two cycles being recorded. Move the scan rate slider and the whole simulation is recalculated. Switch between our ZP 501 carbon electrode and an ideal fast electrode, and tick Show measured data to compare the simulation with a real measurement on a ZP 501 electrode.
Scan rate 50 mV s⁻¹ · run time 56 s · cycle 2: oxidation peak +0.345 V (1.6 µA), reduction peak −0.134 V (−3.4 µA), ΔEp = 480 mV
Simulation: 5 mM ferricyanide only, one electron, D = 7.2 × 10⁻⁶ cm² s⁻¹, E°′ = +0.11 V, α = 0.5, electroactive area 0.0064 cm², planar diffusion with Butler–Volmer kinetics plus a small charging current. The slow-electrode rate constant (k⁰ = 5 × 10⁻⁵ cm s⁻¹) was chosen to match the measured peak separation. The animation is sped up; the time readout shows real time.
What to look for
- Faster scans give bigger peaks. The peak current grows with the square root of the scan rate – the straight line in the lower chart. A faster sweep gives the solution less time to build a thick, depleted layer next to the electrode, so more molecules reach the surface each second.
- On a slow electrode, the peaks move apart as you scan faster. Electron transfer on untreated carbon takes time. Sweep faster, and the potential has to travel further before the reaction keeps up – so the oxidation peak shifts positive, the reduction peak shifts negative, and the gap between them grows.
- On an ideal fast electrode, the peaks stay put – about 60 mV apart whatever the scan rate. That fixed gap is the signature of a reversible one-electron reaction.
- Very fast scans need a wider window. On the slow electrode above about 300 mV s⁻¹, the reduction peak is pushed towards the −0.2 V end of the sweep. To see it in full, you would extend the potential range.
The equation behind the straight line
For a reaction controlled by diffusion, the peak current follows the Randles–Ševčík equation. For a fast, reversible reaction at 25 °C:
ip = 2.69 × 10⁵ · n^3/2 · A · D^½ · C · v^½
where n is the number of electrons, A the electrode area (cm²), D the diffusion coefficient (cm² s⁻¹), C the concentration (mol cm⁻³) and v the scan rate (V s⁻¹). For a slow electrode like untreated carbon, a closely related form applies. Either way, plotting peak current against √v gives a straight line – and its slope tells you the electrochemically active area of your electrode.
In practice, a scan-rate series is one of the most useful experiments you can run on a new electrode: a handful of CVs, each taking about a minute on a screen-printed electrode, tells you whether the reaction is diffusion-controlled, how fast the electron transfer is, and how much of the surface is really working.
Try it yourself
Everything in this simulator can be measured for real with a potentiostat and a screen-printed electrode. A single drop of sample on the electrode is enough to run a two-cycle CV in about a minute.
SenseitAll (SIA) potentiostatA hardware platform for teaching, research and prototyping your own sensor ideas.View the instrument →
Screen-printed electrodesCarbon, gold, platinum and silver/silver chloride electrodes, including the ZP 501 carbon electrode.Browse electrodes →Show the numbers behind the simulation
| Scan rate | ZP 501 carbon: ΔEp | ZP 501 carbon: reduction peak | Ideal fast electrode: ΔEp | Ideal fast electrode: reduction peak |
|---|---|---|---|---|
| 10 mV s⁻¹ | 398 mV | −1.5 µA | 62 mV | −2.0 µA |
| 20 mV s⁻¹ | 434 mV | −2.1 µA | 62 mV | −2.9 µA |
| 50 mV s⁻¹ | 480 mV | −3.4 µA | 62 mV | −4.5 µA |
| 100 mV s⁻¹ | 516 mV | −4.8 µA | 62 mV | −6.4 µA |
| 200 mV s⁻¹ | 552 mV | −6.9 µA | 62 mV | −9.2 µA |
| 500 mV s⁻¹ | 597* mV | −11.1* µA | 62 mV | −14.6 µA |
Cycle 2 values from the simulation. *At 500 mV s⁻¹ the reduction peak on the slow electrode is at the edge of the −0.2 V window, so it is not fully developed.
Model: 5 mM ferricyanide only, one electron, D = 7.2 × 10⁻⁶ cm² s⁻¹, E°′ = +0.11 V, α = 0.5, electroactive area 0.0064 cm², planar diffusion with Butler–Volmer kinetics and a small charging current. ZP 501 carbon: k⁰ = 5 × 10⁻⁵ cm s⁻¹, chosen to match a measured peak separation of about 480 mV at 50 mV s⁻¹. Ideal fast electrode: k⁰ = 1 cm s⁻¹.